Home » GAS MANIFOLD MC80

GAS MANIFOLD MC80

Whole medical range is available on

Manifold MC80 HP UNIT is a fi rst stage pressure regulator intended for use in hospital pipeline systém as a medical gas source.

MC80 HP UNIT contains two regulators with pressure relieve valves and it is connected to two various cylinder banks using high pressure hoses. When the cylinder bank, which has been connected for operation, has been emptied the other duty side is automatically connected.

It is equipped with pressure switches and pressure transmitters 4 – 20 mA. Together with MC80 HP UNIT shall always be used gas alarm providing all alarms in compliance with the standard EN ISO 7396-1.

SPECIFICATION

MC80 HP UNIT INCLUDES THE FOLLOWING COMPONENTS:

  • MC80 HP Manifold

FOR A COMPLETE MC80 HP UNIT MANIFOLD ADD:

  • Gas alarm (optional accessories)
  • High pressure hoses (optional accessories)
Item No.  Denomination Gas
PRV
Alarm
0727321 MC80 HP 2x1 O2 With pull-ring switch + 4–20 mA
0727322 MC80 HP 2x1 Air With pull-ring switch + 4–20 mA
0727323 MC80 HP 2x1 N2O/O With pull-ring switch + 4-20 mA

735100000867 MC80 leaflet EN

TECHNICAL DATA

Gases: O2, Air, N2, N2O, CO2,  (all medical gases)
Nominal flow: 200 m3/h
Inlet maximal pressure: 200 bar (20 000 kPa)
Outlet nominal pressure: 12 bar (setting range 10–16 bar)
Inlet connection: W21,8×1/14“M
Outlet connection: G3/4"F
Pressure relieve valve: 17 bar
Pressure relieve valve
pipe dimension:

∅ 10 mm

Purge valves connection: W21,8×1/14“M + pipe ø 15 mm

Regulatory status:

Complies with Medical Devices Directive 93/42/EEC

Complies with EN ISO 7396-1
(Central Gas Supply Systems)

Complies with EN 60601-1-2
(Electromagnetic compatibility)

present SIS HB 370 and HTM 02-01

 